Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2331A (Expiry date: 24.05.2025)

 

Production date : 25.09.2023

Results of quality control 3 (after 12 months), date: 19.09.2024

No serum blank (average) : 0.121

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.001 -0.003 0.004 0.000 0
Negative control serum
(N° 61
00-06)
0.044 0.038 0.042 0.048 0.043 2
Weak positive serum
(N° 61
00-07)
0.652 0.699 0.618 0.648 0.655 30
Positive control serum
(N° 61
00-08)
2.098 2.176 2.196 2.324 2.199 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.199 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 30 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.121 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch