#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2331A (Expiry date: 24.05.2025)

 

Production date : 25.09.2023

Results of quality control 2 (after 6 months), date: 22.03.2024

No serum blank (average) : 0.137

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.014 0.003 -0.007 0.018 0.000 0
Negative control serum
(N° 61
00-06)
0.037 0.086 0.044 0.111 0.069 3
Weak positive serum
(N° 61
00-07)
0.680 0.650 0.587 0.702 0.654 29
Positive control serum
(N° 61
00-08)
2.206 2.184 2.167 2.342 2.224 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.224 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 29 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.137 Yes

Comments:

Good for use.