#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2331A (Expiry date: 24.05.2025)

 

Production date : 25.09.2023

Results of quality control 1 (after production), date: 28.09.2023

No serum blank (average) : 0.120

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.001 0.003 0.001 0.000 0
Negative control serum
(N° 61
00-06)
0.055 0.086 0.057 0.075 0.069 3
Weak positive serum
(N° 61
00-07)
0.610 0.623 0.524 0.563 0.580 27
Positive control serum
(N° 61
00-08)
2.136 2.182 2.111 2.165 2.149 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.149 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 27 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.120 Yes

Comments:

Good for use.