#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2330M (Expiry date: 24.05.2025)
Production date : 25.09.2023
Results of quality control 3 (after
12 months), date: 19.09.2024
No serum blank (average) : 0.129
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.003 |
0.003 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.021 |
0.035 |
0.032 |
0.025 |
0.028 |
1 |
Weak positive serum
(N° 9300-07) |
0.785 |
0.776 |
0.805 |
0.785 |
0.787 |
31 |
Positive control serum
(N° 9300-08) |
2.413 |
2.563 |
2.451 |
2.638 |
2.516 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.516 |
Yes |
Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
31 % |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.129 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|