#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2330M (Expiry date: 24.05.2025)

 

Production date : 25.09.2023

Results of quality control 3 (after 12 months), date: 19.09.2024

No serum blank (average) : 0.129

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.003 0.003 0.003 0.000 0
Negative control serum
(N° 9300-06)
0.021 0.035 0.032 0.025 0.028 1
Weak positive serum
(N° 9300-07)
0.785 0.776 0.805 0.785 0.787 31
Positive control serum
(N° 9300-08)
2.413 2.563 2.451 2.638 2.516 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.516 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 31 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.129 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland