#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2330M (Expiry date: 24.05.2025)
Production date : 25.09.2023
Results of quality control 2 (after
6 months), date: 22.03.2024
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.014 |
0.009 |
-0.006 |
0.011 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.007 |
0.051 |
0.007 |
0.056 |
0.030 |
1 |
Weak positive serum
(N° 9300-07) |
0.720 |
0.713 |
0.763 |
0.756 |
0.738 |
30 |
Positive control serum
(N° 9300-08) |
2.323 |
2.422 |
2.473 |
2.529 |
2.437 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.437 |
Yes |
Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
30 % |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|