#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2330M (Expiry date: 24.05.2025)

 

Production date : 25.09.2023

Results of quality control 1 (after production), date: 28.09.2023

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.004 0.006 0.002 0.000 0
Negative control serum
(N° 9300-06)
0.011 0.038 0.022 0.030 0.025 1
Weak positive serum
(N° 9300-07)
0.735 0.663 0.698 0.697 0.698 30
Positive control serum
(N° 9300-08)
2.316 2.303 2.379 2.436 2.358 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.358 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 30 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.