Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2329C (Expiry date: 24.05.2025)

Production date : 25.09.2023

Results of quality control 3 (after 12 months), date: 19.09.2024

No serum blank (average) : 0.192

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.004 0.004 0.002 0.000 0
Negative control serum
(N° 9700-06)
0.100 0.104 0.115 0.113 0.108 4
Weak positive serum
(N° 9700-07)
0.472 0.430 0.451 0.481 0.458 19
Positive control serum
(N° 9700-08)
2.433 2.498 2.422 2.542 2.473 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.473 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 19 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.192 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland