Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2329C (Expiry date: 24.05.2025)

Production date : 25.09.2023

Results of quality control 1 (after production), date: 27.09.2023

No serum blank (average) : 0.215

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.008 0.005 0.003 0.000 0
Negative control serum
(N° 9700-06)
0.059 0.097 0.077 0.103 0.084 4
Weak positive serum
(N° 9700-07)
0.446 0.438 0.449 0.448 0.446 19
Positive control serum
(N° 9700-08)
2.301 2.376 2.353 2.492 2.381 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.381 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 19 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.215 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland