#9800
Anisakidae IgG ELISA
ELISA kit for the diagnosis of human anisakidosis
Lot 2328N (Expiry date: 03.05.2025)
Production date : 04.09.2023
Results of quality control 1 (after
production), date: 07.09.2023
No serum blank (average) : 0.128
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.006 |
-0.003 |
0.007 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9800-06) |
0.081 |
0.086 |
0.055 |
0.068 |
0.072 |
3 |
Weak positive serum
(N° 9800-07) |
0.388 |
0.363 |
0.390 |
0.378 |
0.379 |
17 |
Positive control serum
(N° 9800-08) |
2.180 |
2.138 |
2.188 |
2.357 |
2.215 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9800-08 |
> 1.200 |
2.215 |
Yes |
Absorbance of weak positive serum 9800-07, as % of positive serum |
> 9 % |
17 % |
Yes |
Absorbance of negative serum 9800-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.128 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|