#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2326F (Expiry date: 03.05.2025)

Production date : 04.09.2023

Results of quality control 4 (after 20 months), date: 05.05.2025

No serum blank (average) : 0.122

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.003 0.003 0.004 0.000 0
Negative control serum
(N° 9310-06)
0.010 0.013 0.011 0.012 0.012 0
Weak positive serum
(N° 9310-07)
0.126 0.122 0.125 0.138 0.128 5
Positive control serum
(N° 9310-08)
2.304 2.413 2.292 2.402 2.353 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.353 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 5 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 0 % Yes
Absorbance of no serum blank < 0.350 0.122 Yes

Comments:

Good for use.