#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2325G (Expiry date: 03.05.2025)
Production date : 04.09.2023
Results of quality control 4 (after
20 months), date: 05.05.2025
No serum blank (average) : 0.106
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.001 |
0.002 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.054 |
0.052 |
0.067 |
0.060 |
0.058 |
2 |
Weak positive serum
(N° 9350-07) |
1.204 |
1.190 |
1.211 |
1.182 |
1.196 |
52 |
Positive control serum
(N° 9350-08) |
2.267 |
2.299 |
2.282 |
2.406 |
2.313 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.313 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
52 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.106 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|