#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2325G (Expiry date: 03.05.2025)
Production date : 04.09.2023
Results of quality control 3 (after
12 months), date: 22.08.2024
No serum blank (average) : 0.100
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.004 |
0.002 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.070 |
0.074 |
0.077 |
0.079 |
0.075 |
3 |
Weak positive serum
(N° 9350-07) |
1.214 |
1.222 |
1.180 |
1.166 |
1.196 |
51 |
Positive control serum
(N° 9350-08) |
2.262 |
2.451 |
2.292 |
2.382 |
2.347 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.347 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
51 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.100 |
Yes |
Comments:
Good for use.