#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2325G (Expiry date: 03.05.2025)
Production date : 04.09.2023
Results of quality control 2 (after
6 months), date: 28.02.2024
No serum blank (average) : 0.141
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.029 |
0.037 |
-0.033 |
0.025 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.027 |
0.131 |
0.030 |
0.117 |
0.076 |
3 |
Weak positive serum
(N° 9350-07) |
1.100 |
1.175 |
1.115 |
1.126 |
1.129 |
49 |
Positive control serum
(N° 9350-08) |
2.237 |
2.411 |
2.164 |
2.352 |
2.291 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.291 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
49 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.141 |
Yes |
Comments:
Good for use.