#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2325G (Expiry date: 03.05.2025)
Production date : 04.09.2023
Results of quality control 1 (after
production), date: 08.09.2023
No serum blank (average) : 0.113
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.001 |
0.000 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.065 |
0.085 |
0.058 |
0.087 |
0.074 |
3 |
Weak positive serum
(N° 9350-07) |
1.242 |
1.289 |
1.264 |
1.238 |
1.258 |
52 |
Positive control serum
(N° 9350-08) |
2.299 |
2.504 |
2.342 |
2.458 |
2.401 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.401 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
52 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.