#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2323T (Expiry date: 12.03.2025)
Production date : 13.07.2023
Results of quality control 3 (after
12 months), date: 09.07.2024
No serum blank (average) : 0.136
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.023 |
0.071 |
-0.022 |
-0.025 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
-0.003 |
-0.001 |
0.011 |
-0.001 |
0.001 |
0 |
Weak positive serum
(N° 9200-07) |
1.553 |
1.461 |
1.456 |
1.405 |
1.469 |
65 |
Positive control serum
(N° 9200-08) |
2.229 |
2.364 |
2.202 |
2.232 |
2.257 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.257 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
65 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.136 |
Yes |
Comments:
Good for use.