#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2323T (Expiry date: 12.03.2025)

 

Production date : 13.07.2023

Results of quality control 3 (after 12 months), date: 09.07.2024

No serum blank (average) : 0.136

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.023 0.071 -0.022 -0.025 0.000 0
Negative control serum
(N° 9200-06)
-0.003 -0.001 0.011 -0.001 0.001 0
Weak positive serum
(N° 9200-07)
1.553 1.461 1.456 1.405 1.469 65
Positive control serum
(N° 9200-08)
2.229 2.364 2.202 2.232 2.257 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.257 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 65 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 0 % Yes
Absorbance of no serum blank < 0.350 0.136 Yes

Comments:

Good for use.