#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2323T (Expiry date: 12.03.2025)
Production date : 13.07.2023
Results of quality control 2 (after
6 months), date: 10.01.2024
No serum blank (average) : 0.119
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.008 |
0.011 |
-0.002 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.019 |
0.063 |
0.034 |
0.048 |
0.041 |
2 |
Weak positive serum
(N° 9200-07) |
1.519 |
1.582 |
1.583 |
1.598 |
1.570 |
66 |
Positive control serum
(N° 9200-08) |
2.286 |
2.468 |
2.301 |
2.484 |
2.384 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.384 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
66 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.119 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|