#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2322R (Expiry date: 12.03.2025)
Production date : 13.07.2023
Results of quality control 4 (after
20 months), date: 13.03.2025
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.007 |
-0.004 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.052 |
0.053 |
0.053 |
0.061 |
0.054 |
2 |
Weak positive serum
(N° 9750-07) |
0.453 |
0.424 |
0.429 |
0.486 |
0.448 |
19 |
Positive control serum
(N° 9750-08) |
2.299 |
2.302 |
2.321 |
2.483 |
2.351 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.351 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
19 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|