#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2321B (Expiry date: 05.03.2025)

 

Production date : 06.07.2023

Results of quality control 1 (after production), date: 11.07.2023

No serum blank (average) : 0.137

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 0.001 -0.002 -0.001 0.000 0
Negative control serum
(N° 9600-06)
0.067 0.060 0.056 0.069 0.063 3
Weak positive serum
(N° 9600-07)
0.484 0.469 0.506 0.499 0.489 21
Positive control serum
(N° 9600-08)
2.291 2.284 2.294 2.386 2.313 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.313 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 21 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.137 Yes

Comments:

Good for use.