#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2320S (Expiry date : 05.03.2025)
Production date : 06.07.2023
Results of quality control 3 (after
12 months), date: 28.06.2024
No serum blank (average) : 0.115
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.006 |
-0.006 |
0.004 |
0.007 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.083 |
0.090 |
0.082 |
0.098 |
0.089 |
4 |
Weak positive serum
(N° 9450-07) |
0.737 |
0.757 |
0.733 |
0.798 |
0.757 |
36 |
Positive control serum
(N° 9450-08) |
2.045 |
2.137 |
2.098 |
2.111 |
2.098 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.098 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.115 |
Yes |
Comments:
Good for use.