Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2320S (Expiry date : 05.03.2025)

Production date : 06.07.2023

Results of quality control 2 (after 6 months), date: 05.01.2024

No serum blank (average) : 0.142

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.009 0.008 -0.003 0.005 0.000 0
Negative control serum
(N° 9450-06)
0.092 0.145 0.087 0.126 0.112 5
Weak positive serum
(N° 9450-07)
0.798 0.730 0.747 0.732 0.752 34
Positive control serum
(N° 9450-08)
2.141 2.225 2.152 2.232 2.187 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.187 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 34 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.142 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bâtiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch