#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2319S (Expiry date : 11.02.2025)

Production date : 12.06.2023

Results of quality control 3 (after 12 months), date: 06.06.2024

No serum blank (average) : 0.114

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.013 -0.005 -0.003 -0.005 0.000 0
Negative control serum
(N° 9450-06)
0.086 0.071 0.078 0.072 0.077 4
Weak positive serum
(N° 9450-07)
0.709 0.652 0.668 0.677 0.677 33
Positive control serum
(N° 9450-08)
2.010 2.054 1.964 2.099 2.032 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.032 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 33 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.114 Yes

Comments:

Good for use.