#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2319S (Expiry date : 11.02.2025)
Production date : 12.06.2023
Results of quality control 2 (after
6 months), date: 07.12.2023
No serum blank (average) : 0.142
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.015 |
0.020 |
-0.023 |
0.017 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.090 |
0.145 |
0.059 |
0.153 |
0.112 |
5 |
Weak positive serum
(N° 9450-07) |
0.806 |
0.822 |
0.763 |
0.795 |
0.797 |
36 |
Positive control serum
(N° 9450-08) |
2.130 |
2.344 |
2.079 |
2.312 |
2.217 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.217 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.142 |
Yes |