Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2318H (Expiry date: 21.01.2025)

 

Production date : 22.05.2023

Results of quality control 3 (after 12 months), date: 16.05.2024

No serum blank (average) : 0.117

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.003 0.004 -0.003 0.000 0
Negative control serum
(N° 9550-06)
0.169 0.154 0.167 0.149 0.159 7
Weak positive serum
(N° 9550-07)
0.626 0.634 0.656 0.649 0.641 27
Positive control serum
(N° 9550-08)
2.389 2.442 2.344 2.195 2.342 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.342 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  12 % 27 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 7 % Yes
Absorbance of no serum blank < 0.350 0.117 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland