#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2318H (Expiry date: 21.01.2025)

 

Production date : 22.05.2023

Results of quality control 2 (after 6 months), date: 16.11.2023

No serum blank (average) : 0.119

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.008 -0.007 0.005 -0.007 0.000 0
Negative control serum
(N° 9550-06)
0.155 0.161 0.158 0.155 0.158 7
Weak positive serum
(N° 9550-07)
0.585 0.609 0.530 0.551 0.569 25
Positive control serum
(N° 9550-08)
2.319 2.437 2.153 2.258 2.292 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.292 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  12 % 25 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 7 % Yes
Absorbance of no serum blank < 0.350 0.119 Yes

Comments:

Good for use.