#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2317V (Expiry date: 21.01.2025)

 

Production date : 22.05.2023

Results of quality control 2 (after 6 months), date: 16.11.2023

No serum blank (average) : 0.115

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.004 -0.001 0.000 0.000 0
Negative control serum
(N° 9400-06)
0.101 0.150 0.114 0.129 0.124 5
Weak positive serum
(N° 9400-07)
0.930 1.029 0.951 0.988 0.975 40
Positive control serum
(N° 9400-08)
2.261 2.635 2.315 2.560 2.443 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.443 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 40 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.115 Yes

Comments:

Good for use.