#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2317V (Expiry date: 21.01.2025)

 

Production date : 22.05.2023

Results of quality control 1 (after production), date: 26.05.2023

No serum blank (average) : 0.119

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.001 0.002 0.004 0.000 0
Negative control serum
(N° 9400-06)
0.109 0.119 0.128 0.134 0.122 5
Weak positive serum
(N° 9400-07)
1.016 1.013 1.056 0.963 1.012 43
Positive control serum
(N° 9400-08)
2.208 2.355 2.314 2.430 2.327 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.359 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 43 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.119 Yes

Comments:

Good for use.