#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2315A (Expiry date: 02.12.2024)

 

Production date : 03.04.2023

Results of quality control 3 (after 12 months), date: 27.03.2024

No serum blank (average) : 0.129

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.014 0.018 -0.016 0.012 0.000 0
Negative control serum
(N° 61
00-06)
0.028 0.079 0.053 0.090 0.063 3
Weak positive serum
(N° 61
00-07)
0.532 0.561 0.501 0.508 0.526 24
Positive control serum
(N° 61
00-08)
2.151 2.234 2.194 2.304 2.221 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.221 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 24 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.129 Yes

Comments:

Good for use.