#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2315A (Expiry date: 02.12.2024)

 

Production date : 03.04.2023

Results of quality control 2 (after 6 months), date: 29.09.2023

No serum blank (average) : 0.114

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.002 0.001 0.000 0.000 0
Negative control serum
(N° 61
00-06)
0.072 0.077 0.061 0.060 0.068 3
Weak positive serum
(N° 61
00-07)
0.566 0.591 0.528 0.550 0.559 25
Positive control serum
(N° 61
00-08)
2.179 2.361 2.192 2.354 2.272 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.272 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 25 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.114 Yes

Comments:

Good for use.