#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2315A (Expiry date: 02.12.2024)

 

Production date : 03.04.2023

Results of quality control 1 (after production), date: 06.04.2023

No serum blank (average) : 0.114

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.003 0.003 0.002 0.000 0
Negative control serum
(N° 61
00-06)
0.050 0.047 0.066 0.064 0.057 3
Weak positive serum
(N° 61
00-07)
0.455 0.458 0.444 0.467 0.456 22
Positive control serum
(N° 61
00-08)
2.058 2.168 1.994 2.103 2.081 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.081 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 22 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.114 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland