#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2314F (Expiry date: 02.12.2024)

Production date : 03.04.2023

Results of quality control 3 (after 12 months), date: 27.03.2024

No serum blank (average) : 0.125

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 0.012 -0.010 0.004 0.000 0
Negative control serum
(N° 9310-06)
0.016 0.055 0.009 0.055 0.033 2
Weak positive serum
(N° 9310-07)
0.088 0.133 0.089 0.142 0.113 6
Positive control serum
(N° 9310-08)
1.914 2.084 1.971 2.167 2.034 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.034 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 6 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 2 % Yes
Absorbance of no serum blank < 0.350 0.125 Yes

Comments:

Good for use.