Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2314F (Expiry date: 02.12.2024)

Production date : 03.04.2023

Results of quality control 2 (after 6 months), date: 29.09.2023

No serum blank (average) : 0.120

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 -0.005 0.002 0.002 0.000 0
Negative control serum
(N° 9310-06)
0.015 0.032 0.016 0.028 0.022 1
Weak positive serum
(N° 9310-07)
0.104 0.127 0.106 0.131 0.117 5
Positive control serum
(N° 9310-08)
2.107 2.370 2.138 2.257 2.218 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.218 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 5 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.120 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland