#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2314F (Expiry date: 02.12.2024)

Production date : 03.04.2023

Results of quality control 1 (after production), date: 06.04.2023

No serum blank (average) : 0.119

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.002 0.002 -0.002 0.000 0
Negative control serum
(N° 9310-06)
0.025 0.036 0.028 0.033 0.030 1
Weak positive serum
(N° 9310-07)
0.126 0.135 0.132 0.140 0.133 6
Positive control serum
(N° 9310-08)
2.157 2.243 2.065 2.360 2.206 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.206 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 6 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.119 Yes

Comments:

Good for use.