#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2312G (Expiry date: 15.11.2024)
Production date : 16.03.2023
Results of quality control 4 (after
20 months), date: 19.11.2024
No serum blank (average) : 0.133
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
-0.007 |
0.000 |
0.012 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.039 |
0.044 |
0.050 |
0.049 |
0.046 |
2 |
Weak positive serum
(N° 9350-07) |
1.115 |
1.109 |
1.219 |
1.269 |
1.178 |
52 |
Positive control serum
(N° 9350-08) |
2.087 |
2.186 |
2.253 |
2.509 |
2.259 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.259 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
52 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.