#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2312G (Expiry date: 15.11.2024)
Production date : 16.03.2023
Results of quality control 3 (after
12 months), date: 11.03.2024
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.011 |
0.015 |
-0.010 |
0.007 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.041 |
0.103 |
0.036 |
0.083 |
0.066 |
3 |
Weak positive serum
(N° 9350-07) |
1.194 |
1.116 |
1.039 |
1.127 |
1.119 |
52 |
Positive control serum
(N° 9350-08) |
2.047 |
2.322 |
2.076 |
2.207 |
2.163 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.163 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
52 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.