#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2312G (Expiry date: 15.11.2024)
Production date : 16.03.2023
Results of quality control 2 (after
6 months), date: 14.09.2023
No serum blank (average) : 0.119
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.003 |
-0.001 |
-0.001 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.056 |
0.054 |
0.060 |
0.058 |
0.057 |
2 |
Weak positive serum
(N° 9350-07) |
1.144 |
1.204 |
1.193 |
1.201 |
1.185 |
52 |
Positive control serum
(N° 9350-08) |
2.251 |
2.199 |
2.328 |
2.339 |
2.279 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.279 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
52 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.119 |
Yes |
Comments:
Good for use.