#9350 Echinococcus granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses

Lot 2312G (Expiry date: 15.11.2024)

 

Production date : 16.03.2023

Results of quality control 2 (after 6 months), date: 14.09.2023

No serum blank (average) : 0.119

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.003 -0.001 -0.001 -0.002 0.000 0
Negative control serum
(N° 9350-06)
0.056 0.054 0.060 0.058 0.057 2
Weak positive serum
(N° 9350-07)
1.144 1.204 1.193 1.201 1.185 52
Positive control serum
(N° 9350-08)
2.251 2.199 2.328 2.339 2.279 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9350-08 > 1.200 2.279 Yes
Absorbance of weak positive serum 9350-07, as % of positive serum >  27 % 52 % Yes
Absorbance of negative serum 9350-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.119 Yes

Comments:

Good for use.