#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2312G (Expiry date: 15.11.2024)
Production date : 16.03.2023
Results of quality control 1 (after
production), date: 21.03.2023
No serum blank (average) : 0.120
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.004 |
0.006 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.064 |
0.069 |
0.065 |
0.069 |
0.066 |
3 |
Weak positive serum
(N° 9350-07) |
1.313 |
1.284 |
1.299 |
1.361 |
1.314 |
53 |
Positive control serum
(N° 9350-08) |
2.389 |
2.560 |
2.340 |
2.614 |
2.475 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.475 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
53 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.120 |
Yes |
Comments:
Good for use.