Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2308H (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 3 (after 12 months), date: 13.02.2024

No serum blank (average) : 0.128

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.011 0.007 -0.002 0.006 0.000 0
Negative control serum
(N° 9550-06)
0.171 0.227 0.179 0.229 0.201 8
Weak positive serum
(N° 9550-07)
0.597 0.596 0.728 0.784 0.676 27
Positive control serum
(N° 9550-08)
2.241 2.324 2.711 2.851 2.531 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.531 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  12 % 27 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 8 % Yes
Absorbance of no serum blank < 0.350 0.128 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland