#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2307M (Expiry date: 14.10.2024)
Production date : 15.02.2023
Results of quality control 4 (after
20 months), date: 15.10.2024
No serum blank (average) : 0.137
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.007 |
0.003 |
-0.005 |
0.010 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.022 |
0.011 |
0.028 |
0.024 |
0.021 |
1 |
Weak positive serum
(N° 9300-07) |
0.764 |
0.746 |
0.768 |
0.782 |
0.765 |
30 |
Positive control serum
(N° 9300-08) |
2.423 |
2.652 |
2.524 |
2.680 |
2.570 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.570 |
Yes |
Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
30 % |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.137 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|