Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2307M (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 3 (after 12 months), date: 13.02.2024

No serum blank (average) : 0.122

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.004 -0.002 0.007 0.000 0
Negative control serum
(N° 9300-06)
0.013 0.041 0.033 0.076 0.041 2
Weak positive serum
(N° 9300-07)
0.654 0.712 0.663 0.743 0.693 29
Positive control serum
(N° 9300-08)
2.293 2.563 2.285 2.440 2.396 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.396 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 29 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.122 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland