#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2307M (Expiry date: 14.10.2024)
Production date : 15.02.2023
Results of quality control 2 (after
6 months), date: 14.08.2023
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.003 |
-0.001 |
0.005 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.026 |
0.040 |
0.035 |
0.034 |
0.034 |
1 |
Weak positive serum
(N° 9300-07) |
0.754 |
0.772 |
0.739 |
0.755 |
0.755 |
29 |
Positive control serum
(N° 9300-08) |
2.624 |
2.732 |
2.550 |
2.644 |
2.638 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.638 |
Yes |
Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
29 % |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|