Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2307M (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 1 (after production), date: 17.02.2023

No serum blank (average) : 0.129

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.002 0.004 -0.002 0.000 0
Negative control serum
(N° 9300-06)
0.045 0.029 0.045 0.036 0.038 2
Weak positive serum
(N° 9300-07)
0.696 0.660 0.685 0.696 0.684 28
Positive control serum
(N° 9300-08)
2.394 2.456 2.364 2.479 2.423 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.423 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 28 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.129 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch