#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2307M (Expiry date: 14.10.2024)
Production date : 15.02.2023
Results of quality control 1 (after
production), date: 17.02.2023
No serum blank (average) : 0.129
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.002 |
0.004 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.045 |
0.029 |
0.045 |
0.036 |
0.038 |
2 |
Weak positive serum
(N° 9300-07) |
0.696 |
0.660 |
0.685 |
0.696 |
0.684 |
28 |
Positive control serum
(N° 9300-08) |
2.394 |
2.456 |
2.364 |
2.479 |
2.423 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.423 |
Yes |
Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
28 % |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.129 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|