#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2306D (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 4 (after 20 months), date: 15.10.2024

No serum blank (average) : 0.137

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 0.001 -0.011 0.015 0.000 0
Negative control serum
(N° 965
0-06)
0.090 0.081 0.087 0.085 0.086 3
Weak positive serum
(N° 965
0-07)
0.422 0.463 0.398 0.449 0.433 16
Positive control serum
(N° 9650
-08)
2.367 2.880 2.487 2.844 2.645 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.645 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 16 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.137 Yes

Comments:

Good for use.