Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2306D (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 3 (after 12 months), date: 13.02.2024

No serum blank (average) : 0.119

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.000 0.004 -0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.085 0.099 0.098 0.097 0.095 4
Weak positive serum
(N° 965
0-07)
0.398 0.423 0.413 0.463 0.425 18
Positive control serum
(N° 9650
-08)
2.263 2.387 2.347 2.493 2.373 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.373 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.119 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch