#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2306D (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 1 (after production), date: 20.02.2023

No serum blank (average) : 0.123

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.003 0.002 0.005 0.000 0
Negative control serum
(N° 965
0-06)
0.085 0.084 0.091 0.176 0.109 5
Weak positive serum
(N° 965
0-07)
0.396 0.424 0.386 0.381 0.396 17
Positive control serum
(N° 9650
-08)
2.260 2.386 2.372 2.386 2.351 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.351 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 17 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.123 Yes

Comments:

Good for use.