#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2305B (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 4 (after 20 months), date: 15.10.2024

No serum blank (average) : 0.160

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.009 -0.003 0.000 -0.006 0.000 0
Negative control serum
(N° 9600-06)
0.079 0.092 0.053 0.060 0.071 3
Weak positive serum
(N° 9600-07)
0.549 0.537 0.506 0.535 0.532 20
Positive control serum
(N° 9600-08)
2.542 2.639 2.459 2.750 2.598 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.598 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 20 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.160 Yes

Comments:

Good for use.