Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2305B (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 3 (after 12 months), date: 13.02.2024

No serum blank (average) : 0.145

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.003 0.000 -0.005 0.001 0.000 0
Negative control serum
(N° 9600-06)
0.070 0.090 0.076 0.088 0.081 3
Weak positive serum
(N° 9600-07)
0.522 0.513 0.543 0.517 0.524 22
Positive control serum
(N° 9600-08)
2.380 2.423 2.356 2.474 2.409 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.409 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 22 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.145 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch