#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2305B (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 2 (after 6 months), date: 14.08.2023

No serum blank (average) : 0.169

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 0.020 -0.015 0.001 0.000 0
Negative control serum
(N° 9600-06)
0.082 0.167 0.074 0.110 0.108 4
Weak positive serum
(N° 9600-07)
0.634 0.650 0.572 0.630 0.622 23
Positive control serum
(N° 9600-08)
2.689 2.699 2.729 2.858 2.744 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.744 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 23 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.169 Yes

Comments:

Good for use.