#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2305B (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 1 (after production), date: 21.02.2023

No serum blank (average) : 0.148

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 -0.006 0.003 0.007 0.000 0
Negative control serum
(N° 9600-06)
0.057 0.055 0.081 0.071 0.066 3
Weak positive serum
(N° 9600-07)
0.503 0.509 0.496 0.461 0.493 21
Positive control serum
(N° 9600-08)
2.354 2.390 2.414 2.328 2.372 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.372 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 21 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.148 Yes

Comments:

Good for use.