#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2304T (Expiry date: 24.09.2024)
Production date : 25.01.2023
Results of quality control 4 (after
20 months), date: 25.09.2024
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.002 |
-0.001 |
0.007 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.008 |
0.006 |
0.023 |
0.018 |
0.014 |
1 |
Weak positive serum
(N° 9200-07) |
1.426 |
1.424 |
1.535 |
1.559 |
1.486 |
64 |
Positive control serum
(N° 9200-08) |
2.221 |
2.326 |
2.319 |
2.424 |
2.323 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.323 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
64 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|