#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2304T (Expiry date: 24.09.2024)
Production date : 25.01.2023
Results of quality control 3 (after
12 months), date: 19.01.2024
No serum blank (average) : 0.115
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
0.000 |
0.001 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.014 |
0.018 |
0.019 |
0.039 |
0.023 |
1 |
Weak positive serum
(N° 9200-07) |
1.544 |
1.481 |
1.461 |
1.442 |
1.482 |
64 |
Positive control serum
(N° 9200-08) |
2.349 |
2.328 |
2.301 |
2.237 |
2.304 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.304 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
64 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.115 |
Yes |
Comments:
Good for use.